Location
Miami Florida 33166 USA
Phones
(1) 305 704 8548
E-Mail
info@mcscorpusa.com
Free Shipping

Blocks for Magnetic Tests


Blocks for Magnetic Tests

of Novotest

Description

We produce different types of magnetic test saples, such as: bars, plates, disks with different types of cracks.

The test blocks are used for:

  • quality comparison of different magnetic powders (particle solutions);
  • quality assurance of magnetic powders while working with the testing bench;
  • for quality analysis during the magnetic particle testing, also to check the correct dosing of magnetic powder in automatic testing.

MTU Test Block #3
The MTU test block #3 (is the reference block type 1 according to EN ISO 9934-2) is used to verify the indication sensitivity of magnetic particle crack detection agents.
The test block is a disk with two types (large and small) of natural cracks (in both sides).
The large cracks are sanding cracks and the small cracks are tension cracks.
The test block has been magnetized by use of the hole in the middle of the test block. The analys of the test block is made by comparison of the picture with the results of the test block with fluorescent magnetic powder and ultraviolet lamp.

Features

  • Competitive price

Description

We produce different types of magnetic test saples, such as: bars, plates, disks with different types of cracks.

The test blocks are used for:

  • quality comparison of different magnetic powders (particle solutions);
  • quality assurance of magnetic powders while working with the testing bench;
  • for quality analysis during the magnetic particle testing, also to check the correct dosing of magnetic powder in automatic testing.

MTU Test Block #3
The MTU test block #3 (is the reference block type 1 according to EN ISO 9934-2) is used to verify the indication sensitivity of magnetic particle crack detection agents.
The test block is a disk with two types (large and small) of natural cracks (in both sides).
The large cracks are sanding cracks and the small cracks are tension cracks.
The test block has been magnetized by use of the hole in the middle of the test block. The analys of the test block is made by comparison of the picture with the results of the test block with fluorescent magnetic powder and ultraviolet lamp.

  • Competitive price
Novotest-Magnetic-Test-Samples-GrupoTestek-Industry (2)
Specifications
Data Sheet
Buy online
Shape
  • disc
Material
  • steel
Dimensions
  • diameter – 50 mm- thick -11 mm
  • diameter of center hole – 10.5 mm
Preparing of material
  • 90MnCrV8; surfaces grinded on 9,7 ± 0,05 mm
Hardening of material
  • 860 °C to 2 hours- dipped in oil
  • hardness – 63 to 70 HRC
Grinding cracks
  • speedness 35 m/s
  • grain 46J7
  • delivery 0,05 mm cross feed 2,0 mm
Corrosion cracks
  • black heated at 145°C to 150 °C for 1,5 hours
Magnetisation
  • central ladder, direct current I = 1 kA (peek)

 Coming Soon…

Gallery

Subscribe to our contents on topics of importance to the industry

Open chat
1
What do you need?
Hello!

Subscribe to
Download

subscribe to receive promotions and offers, and download the content

You have successfully subscribed to the newsletter

There was an error while trying to send your request. Please try again.

Measurement and Control Solutions Corp will use the information you provide on this form to be in touch with you and to provide updates and marketing.

Apply for information


    By clicking on SEND, you consent to send your data to the companies for which you make a request in order to receive a response to it (Terms and conditions of use).

    Suscribe